Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual

Page 67

Advertising
background image

Directions for Use

Chapter 5 Reviewing Patient Data

63

Range and Alarm Flags

Flag characters (

 and ) indicate a data value that falls outside of a defined range or

violates a programmed alarm threshold. In the illustration above, 69

 indicates a diastolic

pressure value below the programmed alarm limit, and 110.0F

indicates a temperature

value above the monitor’s temperature measurement limit.

Temperature

The temperature notation includes information about the type and location of the
measurement, and can include error notation:

Pulse Rate

The pulse rate notation does not include error information.

SpO

2

SpO

2

notation can include error information:

Type

F

Fahrenheit

C

Celsius

NOR

Normal (predictive)

MON

Monitored

Location

OrL

Oral

(Oral probe)

AP

Axillary Pediatric

(Oral probe)

AA

Axillary Adult

(Oral probe)

rEC

Rectal

(Rectal probe)

Error

(P)

Discontinuity in tissue contact during the measurement
period. The displayed temperature measurement is not
necessarily accurate.

C20

Broken or missing temperature probe.

C7

Malfunctioning sensor.

--

Sensor error.