Chapter 2 Setup

Welch Allyn Vital Signs Monitor 300 Series

Changing the Patient Type

The age range for each patient type is defined as follows:

Default setting: ADULT.

Follow these steps to change the patient type setting.

1.

Press

. The current patient type (

,

, or

) appears below the DIA window, and

NEONATE, PEDIATRIC, or ADULT appears in the message window.

2.

Press

or

to display

,

, or

.

3.

To select the displayed patient type and return to normal operation, either do nothing
for 10 seconds or press any button other than

or

. If you press a function button

(such as

), the monitor returns to normal operation with that function (

) activated.

Changing the patient type has the following effects:

Alarm limits are reset to the default limits for the new patient type

Cuff inflation target pressure is reset to the default for the new patient type

If you cycle through the patient types but do not change the setting, the alarm limits and
the cuff inflation target pressure settings do not change.

Neonatal

Term birth through 28 days, or up to 44 gestational weeks

Pediatric

29 days through 12 years

Adult

13 years and older