A&D Weighing AD-4406 User Manual

Page 14

Advertising
background image

Page 12

AD-4406 Weighing Indicator

Dual Range

<First range> Specify the resolution, decimal point position and format.

ω

<First range> Specify the weighing range.

ω

<Second range> Specify the resolution.

ω

<Second range> Specify the weighing capacity.

For the range function, refer to “5.3. Weighing Range Function”.

5.2.2.

Specifying the Range and Unit

Step1

The range and unit of measure are displayed.

Range display :

SinGL : single range

dUAL : dual range

To change the range function, use the

key

Unit display

The active unit is displayed. The unit can be changed;

such as kg or lb. Calibration is performed using the unit

displayed.

Press the

key to select a unit.

Press the +/- key to select a unit for calibration. The unit for

calibration (first unit) is displayed and the alternate unit (second unit)

is blinking.

ENTER

The key to store the data displayed and proceed to the next step.

ESC

The key to proceed to the next step without changing the data.

5.2.3.

Specifying the Resolution, Decimal Point Position and Format

Step 2 The resolution will be displayed as d 0.1 with decimal point. The indicator displays

triangle 1 and the first unit selected at the previous step.

Specify the position of the decimal point with the

key. Specify the display

format (point or comma) with the +/- key. Specify the resolution with the

1

key.

The decimal point format for serial data output is selected using the F-function

settings.

Press the ENTER key to store the parameters and proceed to the next step.

When pressing the ESC key, regardless of what is displayed, the indicator will

proceed to the next step without changing the parameters.

5.2. Calibration Procedure