AD-4406 Weighing Indicator

Press and hold the ON/OFF key and press the ESC key.

No parameters are changed,

CAnCEL is displayed and the calibration mode is finished.

Step18

Press the ON/OFF key to turn the display off.

5.3.

5.3.

5.3.

5.3. Weighing Range Function

Weighing Range Function

Weighing Range Function

Weighing Range Function

The weighing range function can select "single range" or "dual range". Specify each

weighing interval (division) for the multi-interval instrument. Each weighing interval is

displayed according to a net value or gross value.

Note

When single range is used, performing this function is not required.

Example 1

The gross display.

Specified parameters:

First range

Range = 50.00 kg, division 0.02 kg

Second range

Range = 100.00 kg (maximum capacity), division 0.1 kg

Display

0 kg to 50 kg :

The first range, division 0.02 kg.

50 kg to 100 kg :

The second range, division 0.1 kg.

division 0.02 kg

ޓޓޓ

division 0.1 kg

0 kg 50 kg 100 kg Gross

Weighing value

Example 2

The net display using a 40kg tare value.

Specified parameters:

The same parameters as example 1.

Display

-40 kg to 50 kg :

The first range, division 0.02 kg.

50 kg to 60 kg :

The second range, division 0.1 kg.

division 0.02 kg

division 0.1 kg

-40 kg 0 kg 50 kg 60 kg Net

Tare value

90 kg 100 kg

40 kg Weighing value