AD-4406 Weighing Indicator

12. Hold Function

Hold Function

Hold Function

Hold Function

This function displays the hold weight data after averaging the weight data for a

specific period.

Useful to determine a living animal's weight.

Averaging time is selectable up to 9.9 seconds by 0.1 second step.

3 methods are available to start averaging; manual start, automatic start after

stable, and manual / automatic start.

Manual start is available with key switch or external input.

Serial interface commands are also available; averaging start, releasing the hold

data and outputting the hold state. Refer to “13.3.4. Commands for Hold Function”.

Caution

This function can not be used under the setting CF00=1

Averaging can not start at a displayed value smaller than 0 +/- 5 digits.

Data when the display is over is not included for averaging.

When powered off, the hold is released automatically.

There is no peak hold function.

The display and the Data Output of Hold and Average

The weight display is blinking during the averaging period.

The output data in the averaging period is the actual weight at the time.

The weighing unit is blinking when the weight display is in the hold state.

The output data format of the hold weight data is the same as that of the normal

weight data except the header of stable state is “HD” in the response to the “RW”

or “RW,n” (n=1 or 2) command.

Relations to the Other Functions

If automatic accumulation (F20:21) and/or auto print (F37=2,3 F40=2,3) is set,

accumulation and/or data output is performed after determining the hold data.

12.1. Setting the Hold Functions

F27 determines the method of starting the average.

F27=1

Manual start: Starts the average and release with key switch operation.

F27=2

Automatic start: After passing the inhibit region * and detect stable **,

starts the average automatically, releasing the data when the weight

returns to inhibit region.

F27=3

Both Manual start and Automatic start.

*inhibit region

0 +/- 5digits

**stable detection

Satisfied both F02 and F03

F28 determines the averaging time by 0.1second step. F28=0 holds the data at

averaging start.

12. Hold Function