Zero calibration, Span calibration – A&D Weighing AD-4406 User Manual

Page 16

Advertising
background image

Page 14

AD-4406 Weighing Indicator

F 2

ENTER

F 2

ENTER

Stored input voltage

New input voltage

mV/V

+/

+/

+/

+/-

--

-

ZERO

TARE

mV/V

Place nothing on
the weighing unit

(No load)

Turn on STABLE mark

Zero calibration

To Span Calibration

5.2.8. Zero

Calibration

Step 8

Check the CAL 0 display.

Select a zero calibration method to adjust the zero point.

Weighing input
(Normal way)

The adjustment method
with nothing on the
weighing unit.

To step 9

Digital input

The numerical way to enter
a load cell output voltage.

To step 10

Weighing Input

Step 9

Place nothing on the weighing unit. Press the ENTER

key after the STABLE mark has turned on. The new

zero point parameter will be store. Proceed to step11.

ESC key.......... To proceed to step 11 without changing

the zero point parameter.

Caution Do not press the

ENTER

key while the STABLE

mark is off (detecting motion). Arrange the

condition using the F00 filter function.

Digital Input

Step10

Pressing the

+/-

key, a stored input voltage parameter of

the zero point is displayed in the unit of mV/V. Adjust the

input voltage using the following keys.

key .......... To select a digit.

key .......... To select a value.

ENTER key .... To store the zero point parameter and

proceed to the next step.

ESC key.......... To proceed to the next step without

changing the parameter.

5.2.9. Span

Calibration

Step11

Check the capacity display after CAL F is displayed for 2 seconds.

Select a span calibration method to adjust the capacity.

Weighing a mass less than
the maximum capacity

The method to weigh a mass less
than the maximum capacity.

To step 12

Weighing a maximum
capacity mass

The method to weigh a mass
equivalent to the maximum capacity.

To step 14

Digital input

The numerical way to enter a load cell
output voltage.

To step 16

5.2. Calibration Procedure