Page 22

AD-4406 Weighing Indicator

6.2. F-Functions

F-Functions

F-Functions

F-Functions

Weighing Conditions (Digital filter, Zero trucking and Stability)

Item

Parameter

Description

0

2 d/ 1.6s

1

4 d/ 1.6s

2

8 d/ 1.6s

3

16 d/ 1.6s

4

32 d/ 1.6s

5

64 d/ 1.6s

6

128 d/ 1.6s

7

2 d/ 3.2s

* 8

4 d/ 3.2s

9

8 d/ 3.2s

10

16 d/ 3.2s

11

32 d/ 3.2s

12

64 d/ 3.2s

F00
Filter
Motion / Averaging
time

13

128 d/ 3.2s

If a weak filter is set, the
response will be fast, but will be
more sensitive to external
influences such as vibration.

0

OFF

1

0.5 d/ 1s

2

1.0 d/ 1s

3

1.5 d/ 1s

4

2.0 d/ 1s

5

2.5 d/ 1s

6

0.5 d/ 2s

7

1.0 d/ 2s

* 8

1.5 d/ 2s

9

2.0 d/ 2s

F01
Zero tracking

10

2.5 d/ 2s

This function traces the weight
value drifting around the zero
point slowly, displayed as zero.
If a strong parameter is set, a
very small zero drift may be not
detected.
If CF00 is 1, then 0, 1, 6 or 7
can be selected only for F01.
If CF00 is 1, the initial setting for
F01 is 7.

0

No motion detection

1

0.5 d/ 0.5s

2

1.0 d/ 0.5s

3

2.0 d/ 0.5s

4

3.0 d/ 0.5s

5

4.0 d/ 0.5s

6

0.5 d/ 1s

7

1.0 d/ 1s

* 8

2.0 d/ 1s

9

3.0 d/ 1s

F02
STABLE mark
(Motion detection)
condition

10

4.0 d/ 1s

The function to set the condition
of judgment whether a weigh
value is unstable or stable. The

ZERO key and TARE key are

active in the stable state. If
these keys need to be active in
the unstable state, set to F02 0.
In case of CF00 is 1, F02: 6 or
F02: 7 can be selected only.
(Initial setting is 6.)

0

1 (

STABLE

on)

1

Twice in succession

* 2

Three times

F03
Auto print/Auto
accumulation timing

3

Four times

Set the number of times when
the STABLE mark turned on in
succession, until output. If CF00
is set to 1, F03 2 or 3 can be
selected.

d:

division (weighing interval) of the first range

s: second

*: Initial settings