5.5.

5.5.

5.5.

5.5. Gravity C

Gravity C

Gravity C

Gravity Compensation

ompensation

ompensation

ompensation F

F

F

Function

unction

unction

unction

If the scale is used at the calibration location, it is not necessary to perform this function.

If there is a difference of gravity acceleration between the installed location and

calibration location, it may cause a weighing error. This function specifies these

gravity accelerations and corrects the span error.

Note

The decimal point is not displayed in the function. Example: 9798 = 9.798 m/s

2

When span calibration is executed, the gravity acceleration correction will

be cleared and two gravity acceleration values will return to the factory

settings.

Step 1

At the CAL 0 display, press the

key until G SEt is displayed and press the

ENTER key to enter the gravity compensation function.

If you want to cancel the current procedure, press and hold the ON/OFF key and

press the ESC key. Then, no parameters are changed and the calibration mode is

finished. Press the ON/OFF key to turn the display off after displaying CAnCEL .

Step 2

The parameter is displayed with triangle 1. Enter the gravity acceleration of the

calibration location. The parameter xxxx is the gravity acceleration.

key ...... To select a digit.

key....... To select a value.

ENTER

key .. To store the new gravity acceleration and proceed to step 3.

ESC

key....... To return to G SEt without changing the value.

Step 3

The parameter is displayed with triangle 2. Enter the gravity acceleration of the

installed location. The parameter xxxx is the gravity acceleration.

key ...... To select a digit.

key ...... To select a value.

ENTER key . To store the new gravity acceleration and proceed to step 4.

ESC key ...... To return to step 2 without changing the value.

Step 4

Now G xxxx is displayed. Press the CAL key to store the parameters. CALoFF is

displayed. Proceed to step 5.

Step 5

Press the ON/OFF key to turn the display off.

5.5. Gravity Compensation Function