AD-4406 Weighing Indicator

3.3. The Load Cell Connections

Connect the load cell wires to the connector (receptacle), at the rear panel, using the

accessory load cell plug.

It is possible to connect a 4 wire cable provided that pins 1-2 and pins 3-4 are

shorted, if the distance between the indicator and a load cell is shorter than 5m.

The output voltage of a load cell is a very sensitive signal. Place the load cell cable

away from any noise source.

It is possible to connect four 350 ohm load cells.

The load cell drive is 5VDC 5% between EXC+ and EXC-, the maximum current 60 mA.

3.4. Adjustment of the Load Cell Output

Caution

Use a metal film resistor in the range of 50kohm to 500kohm with a good

temperature coefficient, when adding a resistor to adjust a load cell

output. Use the largest resistance value possible in the range in which

zero adjustment is possible. Solder this resistor at a point near the load

cell or the indicator.

In Case of Reducing the Output Voltage

When the zero output is too large, add a resistor

between EXC+ and SIG-.

In Case of Adding an Offset Voltage to the Output

When the zero output is too small, add a resistor

between EXC+ and SIG+.