AD-4406 Weighing Indicator

5.6.

5.6.

5.6.

5.6. Calibration Error Code List

Calibration Error Code List

Calibration Error Code List

Calibration Error Code List

Exiting from a Calibration Error

ESC key .... To return to the point where an error occurred. Retry the operation.

ESC key while pressing the ON/OFF key.

No parameters are changed,

CAnCEL

is displayed and the calibration

mode is finished.

Error Code List

If an error has occurred during calibration mode, the following code is displayed.

Error code

Description

Err 0

With a multi-interval scale. The last division is set to maximum ( d-50 ).
Therefore the next division can not be entered.

Err 1

The resolution exceeds 40000. (Resolution = maximum capacity/ minimum
division)
Reduce the maximum capacity or increase the minimum division.

Err 2
Err 3

The load cell output is too large or too small at zero calibration. Check the weighing
unit and load cell. Refer to "3.5.Verifying Load Cell Output and Input Sensitivity".

Err 4

Measuring calibration mass (or entering digital input), the value exceeded the
maximum capacity. Reduce calibration mass (or digital input).

Err 5

The selected calibration mass is smaller than the minimum division.

Err 6

The new input sensitivity is less than 0.2 µV/division. Increase the input
sensitivity.
Refer to "3.5.Verifying Load Cell Output and Input Sensitivity".

Err 7

Placing a mass on the weighing unit, the load cell output becomes a negative
value. Check the load cell cable connections and the direction of load cell
mounting.

Err 8

The load cell output exceeds the input range before the maximum capacity.
Adjust the zero balance referring to "3.5.Verifying Load Cell Output and Input
Sensitivity". Replace with a load cell designed for a smaller output. Reduce the
maximum capacity.

Err 9

The weighed value is out of the input range at zero calibration or span
calibration. Check the weighing unit and cables.

Err 12

The previous weighing range is larger than next weighing range.

Err 13

An incorrect mass is selected at the digital linearization function.
Select a mass of the following relation. lnr 1 < lnr 2 < lnr 3.

G Err

An unacceptable value was selected in the gravity acceleration function.