A&D Weighing AD-4406 User Manual

Page 28

Advertising
background image

Page 26

AD-4406 Weighing Indicator

Data Output

Item

Parameter

Description

* 0

No data output

1

Analog output

Set F31, 32, 33

2

Serial output

F30
Data output

3

Serial output (Zero suppressing)

RS-232C,
RS-422/485

*:

Initial settings (In case of installing RS-232C, F30 is set to 2. )

Analog Output

Item

Parameter

Description

* 0

Displayed data

1

Gross data

F31
Output data

2

Net data

F32
Weight value at 4 mA output

-999999 to 999999
(Initial setting is 0)

F33
Weight value at 20 mA output

-999999 to 999999
(Initial setting is 10000)

Polarity is changed
by pressing the +/-
key.

*:

Initial settings

Serial Data Format

Item

Parameter

Description

x 0

Terminator

x 1

Device ID (selected at F06)

x 2

Code number

x 3

Data number

x 4

Result of comparison

x 5

(Reserved)

x 6

(Reserved)

x 7

Accumulated value

x 8

Accumulation count

x 9

Stable/Over

x A

Displayed weight

x B

Gross weight

x C

Net weight

x D

Tare weight

x E

Weight type (G /N / T, refer to CF06)

F34
Serial data format 1

Initial value
19

,2E

,3A,4F,E50

F35
Serial data format 2

Initial value
17,2F,E30

x F

Weight unit (refer to CF07)

100,10:order of
output (1-999)
Select with the

1

1

key.
1: output data
Select with the
key.

"Exxx" indicates
the end of data to
output.

Press NET/GROSS
to expand or shorten
the output data length.

1 “

ጝ” (right hand of the figure) of the display indicates that the comma (F46=0, semi-colon

for F46=1) is output after the data. To turn this mark on and off, press the ZERO and

TARE keys at the same time.

Output of initial settings of F34 (19

ጝ,

2E

, 3A, 4F, 50) is like “ST,GR,12345.6kg”.

2 The data number (parameter 3) increments automatically with each serial data output

from 1 to 99999 (the next of 99999 is 1). The starting number can be set by the keys (F2

+ NET/GROSS).

6.2. F-Functions