AD-4406 Weighing Indicator

Page 13

5.2.4.

Specifying the Weighing Range of the First Range

Step 3 After displaying CAP for 2 seconds, single range or the weighing capacity will be

displayed. When dual range is used, CAP1 is displayed for 2 seconds. Triangle 1

will be displayed. Specify the range or the capacity with the

and keys.

Press the ENTER key to store the parameter and proceed to the next step.

When pressing the ESC key, regardless of what is displayed, the indicator will

proceed to the next step without changing the parameter. The next step is zero

calibration in single range or specifying the second range resolution in dual range.

5.2.5.

Specifying the Second Range Resolution

Step 4 After displaying rAnGE2 for 2 seconds, the resolution with decimal point and

triangle 2 will be displayed. Specify the second range resolution in the same way

as the first range. The decimal point cannot be moved. Specify the second range

resolution greater than the first range.

Press the ENTER key to store the parameter and proceed to the next step.

When pressing the ESC key, regardless of what is displayed, the indicator will

proceed to the next step without changing the parameter.

5.2.6.

Specifying the Second Range Capacity

Step 5

After displaying CAP2 for 2 seconds, the capacity with unit and decimal point will be

displayed. Specify the capacity in the same way as the first range. The capactity

should be greater than the first range.

Press the ENTER key to store the parameter and proceed to zero calibration.

When pressing the ESC key, regardless of what is displayed, the indicator will

proceed to the next step without changing the parameter.

5.2.7.

To Get Stabilized Data

Step 6 Maintain the following conditions to calibrate the scale (indicator) correctly.

Maintain a constant temperature, stable power and stable input voltage from the

load cell.

Avoid direct sunshine or near the outlet of an air conditioner.

Do not install the scale (indicator) where there is a strong magnetic field.

Step 7 Turn the display on and leave it for several minutes.