No stable measured

value

Obviously incorrect

measured values

Cause

Remedy

IDS pH sensor:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if necessary

IDS pH sensor + test sample:

– Conductivity too low

– Use suitable IDS pH sensor

– Temperature too high

– Use suitable IDS pH sensor

– Organic liquids

– Use suitable IDS pH sensor

Cause

Remedy

IDS pH sensor:

– IDS pH sensor unsuitable

– Use suitable IDS sensor

– Temperature difference between

buffer and test sample too great

– Adjust temperature of buffer or

sample solutions

– Measurement procedure not suit-

able

– Follow special procedure