YSI MultiLab 4010-3 User Manual

Page 54

Advertising
background image

Ion concentration

MultiLab 4010-3

54

ba76143e03

08/2013

Calibration

evaluation

After calibrating, the meter automatically evaluates the calibration.

Menu item

Setting/
function

Explanation

Calibration /
Calibration data stor-
age
/Display

-

Displays the calibration records.
Further options:

Scroll through the calibration
records with <><>.

Using <PRT>, output the dis-
played calibration record to the
USB-B (PC) interface or the USB-
A (USB printer) interface.

Using <PRT_>, output all calibra-
tion records to the USB-B (PC)
interface or the USB-A (USB
printer) interface.

Quit the display with <ESC>.

Switch directly to the measured
value display with <M>.

Calibration / Calibra-
tion data storage
/
Output to USB flash
drive or printer

-

Outputs the stored calibration data to
the USB-A interface (USB memory
device/USB printer)

Calibration /
Calibration data stor-
age
/
Output to RS232/USB

-

Outputs the stored calibration data to
the USB-B interface (PC)

Display

Calibration
record

Magnitude of the slope [mV]

+++

50.0 ... 70.0 or 25.0 ... 35.0

Error

Error

< 50 or > 70
or
< 25 or > 35

Error elimination (see section 13 W

HAT

TO

DO

IF

..., page 89)