YSI MultiLab 4010-3 User Manual

Page 95

Advertising
background image

MultiLab 4010-3

Settings

ba76143e03

08/2013

95

Setting menu of

4320

Menu item

Possible
setting

Explanation

Cell constant

0.090 ... ...
0.100 ...
0.110 cm

-1

Display and setting options for the
cell constant

Temp. comp. (TC) /
Method

nLF
lin
off

Procedure for temperature compen-
sation (see section 9.2 T

EMPERA

-

TURE

COMPENSATION

, page 80).

This setting is only available for the
measured parameters,

ϰ and ρ.

Temp. comp. (TC) /
Linear coeff.

0.000 ...
2.000 ...
3.000 %/K

Coefficient of the linear temperature
compensation.
This menu item is only available
when the linear temperature com-
pensation is set.

Temp. comp. (TC) /
Reference temp.

20 °C
25 °C

Reference temperature
This setting is only available for the
measured parameters,

ϰ and ρ.

Multiplier for TDS

0.40 ... 1.00

Factor for TDS value

Reset

-

Resets all sensor settings to the
delivery condition (see section
10.7.1 R

ESETTING

THE

MEASUREMENT

SETTINGS

, page 98).

Advertising