YSI MultiLab 4010-3 User Manual

Page 48

Advertising
background image

Ion concentration

MultiLab 4010-3

48

ba76143e03

08/2013

Stability control

(AutoRead)

& HOLD function

The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display

as soon as the measured value is outside the stability range

when the automatic Stability control is switched off.

You can activate or switch off the automatic Stability control function (see
section 10.6.3 A

UTOMATIC

S

TABILITY

CONTROL

, page 97).

01.08.2013 08:00

1.

Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.

You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.

2.

Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the dis-
play of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.

You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.