3 measuring the temperature, 2 orp calibration, Measuring the temperature – YSI MultiLab 4010-3 User Manual

Page 46: Orp calibration

Advertising
background image

ORP voltage

MultiLab 4010-3

46

ba76143e03

08/2013

6.1.3 Measuring the temperature

For reproducible ORP measurements, it is essential to measure the tempera-
ture of the test sample.
When operating a sensor without integrated temperature sensor, you first have
to measure and enter the temperature of the sample.

The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.

The display of the temperature indicates the active temperature measuring
mode:

6.2

ORP calibration

Temperature
sensor

Resolution of
the temp. dis-
play

Temp. measurement

Yes

0.1 °C

Automatic with temperature sensor

-

1 °C

Manual

ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.

Advertising