MultiLab 4010-3

Ion concentration

ba76143e03

08/2013

51

Continuing with two-

point calibration

9.

Immerse the electrode in standard solution 1.

10.

When calibrating without temperature sensor:

Measure the temperature of the standard solution using a thermom-
eter.

Use <F4>/[

⇅ ] to select the setting of the temperature.

Use <> <> to set the temperature.

Use <F4>/[

⇅ ] to select the setting of the concentration.

11.

Set the concentration of the standard solution with <> <> and
press <ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).

12.

Wait for the end of the AutoRead measurement or accept the calibra-
tion value with <ENTER>.
The calibration display for the next standard solution appears.

01.08.2013 08:00

01.08.2013 08:00

Standard

13.

Thoroughly rinse the electrode with distilled water.

14.

Immerse the electrode in standard solution 2.