YSI MultiLab 4010-3 User Manual

Page 30

Advertising
background image

pH value

MultiLab 4010-3

30

ba76143e03

08/2013

Continuing with two-

point calibration

4.

Thoroughly rinse the sensor with deionized water.

5.

Immerse the sensor in the first buffer solution.

6.

For measurements without temperature sensor
(e.g. when using an IDS adapter

)

:

Enter the temperature of the buffer with <><>.

7.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

8.

Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).

9.

If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.

01.08.2013 08:00

YSI

01.08.2013 08:00

YSI

10.000

10. Thoroughly rinse the sensor with deionized water.

11. Immerse the pH sensor in buffer solution 2.