4 calibration data, Calibration data – YSI MultiLab 4010-3 User Manual

Page 82

Advertising
background image

Conductivity

MultiLab 4010-3

82

ba76143e03

08/2013

9.3.4 Calibration data

The calibration data can be displayed and then output to the interface.

Displaying the cali-

bration record

The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL_> key.

The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.

4.

Immerse the conductivity sensor in the control standard solution,
0.01 mol/l KCI.

5.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

6.

Wait for the end of the measurement with stability control
or
take over the calibrated value with <ENTER>.
The calibration record is displayed and output to the interface.

7.

Switch to the measured value display with <ENTER>.

The calibration record is automatically transmitted to the interface
after calibrating.

Menu item

Setting/
function

Explanation

Calibration /
Calibration data
storage
/Display

-

Displays the calibration records.

Further options:

Scroll through the calibration
records with <><>.

Using <PRT>, output the dis-
played calibration record to the
USB-B (PC) interface or the
USB-A (USB printer) interface .

Using <PRT_>, output all calibra-
tion records to the USB-B (PC)
interface or the USB-A (USB
printer) interface .

Quit the display with <ESC>.

Switch directly to the measured
value display with <M>.