3 selecting the measuring method, Selecting the measuring method – YSI MultiLab 4010-3 User Manual

Page 56

Advertising
background image

Ion concentration

MultiLab 4010-3

56

ba76143e03

08/2013

7.3

Selecting the measuring method

The following methods are supported:

Standard addition

Standard subtraction

Sample addition

Sample subtraction

Blank value addition

1.

Connect the ISE combination electrode to the meter.
The pH/ISE measuring window is displayed.

2.

If necessary, select the ISE display with <M>.

3.

If necessary, measure the temperature of the test sample with a ther-
mometer.

4.

Open the ISE menu with <ENTER>.

5.

Thoroughly rinse the electrode with distilled water.

6.

Adjust the temperature of the standard solutions.

7.

Select Method with <> <> and confirm with <ENTER>.

8.

Select a method with <> <> and confirm with <ENTER>.

9.

Select Start method with <> <> and confirm with <ENTER>.
Measurement with the selected method begins (see section 7.3.1
S

TANDARD

ADDITION

, page 57 ... section 7.3.5 S

TANDARD

ADDITION

WITH

BLANK

VALUE

CORRECTION

(B

LANK

VALUE

ADDITION

), page 66).

Calibration
Man. temperature

:

25 °C

ISE setup
Method

:

Standard addition

Start method

ISE

01.08.2013 08:00