MultiLab 4010-3

pH value

ba76143e03

08/2013

31

Continuing with

three- to five-point

calibration

12. When measuring without temperature sensor:

Enter the temperature of the buffer with <><>.

13. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

14. Wait for the measurement with stability control to be completed or termi-

nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).

15. If necessary, finish the calibration procedure as a two-point calibration

with <M>.
The calibration record is displayed.

01.08.2013 08:00

YSI

16. Thoroughly rinse the sensor with deionized water.

17. Immerse the sensor in the next buffer solution.

18. When measuring without temperature sensor:

Enter the temperature of the buffer with <><>.

19. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.