MultiLab 4010-3

Dissolved oxygen (D.O.)

ba76143e03

08/2013

73

8.2.5 2-point calibration

Calibration in water vapor-saturated air (Cal) and calibration via Comparison
meas.
according to the Winkler method (Comp) are described jointly here.

6.

Wait for the AutoRead measurement to be completed (audio signal).

When calibrating in water vapor-saturated air (Cal):
The calibration record is displayed and output to the interface.

When calibrating via comparison measurement (Comp):
Set the concentration value with <> <> and accept it with
<ENTER>.
The calibration record is displayed and output to the interface.

7.

Switch to the measured value display with <ENTER>.

1.

Connect the D.O. sensor to the meter.

2.

Check the selected calibration type and selected number of calibration
points in the sensor setting menu and, if necessary, change them.

3.

Place the D.O. sensor in a solution that does not contain any dissolved
oxygen.

A solution not containing any dissolved oxygen can be prepared by
dissolving approx. 8 to 10 g sodium sulfite (Na

2

SO

3

) in 500 ml tap-

water. Carefully mix the solution. It may take up to 60 minutes until
the solution is free of oxygen.

4.

Start the calibration with <CAL>.
Cal (calibration in water vapor-saturated air) or Comp (calibration via
Comparison meas. according to the Winkler method) is displayed.
The calibration point 1 (0%) is displayed.

01.08.2013 08:00

760 mmHg

25.1 °C