8 qsc function (sensor quality control), Qsc function (sensor quality control), Ee section 5.2.8 qsc – YSI MultiLab 4010-3 User Manual

Page 40: Function, Sensor, Quality, Control

Advertising
background image

pH value

MultiLab 4010-3

40

ba76143e03

08/2013

5.2.8 QSC function (sensor quality control)

General information

on the QSC function

The QSC function (Quality Sensor Control) is a new sensor evaluation for
digital IDS sensors. It evaluates the condition of an IDS pH sensor individually
and with a very fine grading.

The QSC scale shows the current sensor evaluation with an indicator on the
display.

In the USB output the sensor evaluation is given as a percentage (1-100).

The finely graded sensor evaluation of the QSC function promptly calls your
attention to changes of the sensor.
Thus you can do what is necessary to restore the optimum measuring quality
(e.g. clean, calibrate or replace the sensor).

Sensor evaluation

with / without
QSC function

QSC calibration

The QSC function is enabled by once carrying out an additional three-point cali-
bration with special QSC buffer solutions. It covers the measuring range of the
sensor (pH 2 to pH 11). The QSC initial calibration determines the actual condi-
tion of the sensor and stores it as a reference in the sensor.
To meet the high requirements of a QSC initial calibration, the QSC initial cali-
bration should optimally be carried out with the initial commissioning of the
sensor.

1 QSC scale

The double arrow on the QSC scale indicates the current sensor eval-
uation

01.08.2013 08:00

Info

1

With QSC function

Without QSC function (sensor sym-
bol)

Very fine grading of the sensor eval-
uation (100 grades)

Rough grading of the sensor evalua-
tion (4 grades)

The reference value is individually
determined for each sensor during
the QSC initial calibration.

A theoretical reference value is used
for all sensors

Low tolerances for zero point and
slope when using QSC buffer solu-
tions

Greater tolerances for zero point and
slope when using commercial buffer
sets

Additional QSC calibration required
(with special QSC buffer set)

No additional calibration required

Advertising