Calibration

evaluation

After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.

Calibration record

Calibration / Cali-
bration data stor-
age
/ Output to
USB flash drive or
printer

-

Outputs the stored calibration data
to the USB-A interface (USB mem-
ory device/USB printer)

Calibration /
Calibration data
storage
/ Output to
RS232/USB

-

Outputs the stored calibration data
to the USB-B interface (PC)

Menu item

Setting/
function

Explanation

Display

Calibration record

Cell constant [cm

-1

]

+++

Within the range
0.450 ... 0.500 cm

-1

Error

Error

Outside the range
0.450 ... 0.500 cm

-1

Error elimination (see section 14
W

HAT

TO

DO

IF

..., page 115)

CALIBRATION Cond
03.04.2013 07:43:33

4310
Ser. no. 09250033

Cell constant 0.476 1/cm