1 standard addition, Stand, Section 7.3.1 s – YSI MultiLab 4010-3 User Manual

Page 57: Tandard, Addition

Advertising
background image

MultiLab 4010-3

Ion concentration

ba76143e03

08/2013

57

7.3.1 Standard addition

In the "Standard addition" procedure, a known amount of standard solution is
added to the sample.
The ion concentration in the sample is calculated from the change in potential.

Measuring

1.

Select the measuring method (see section 7.3 S

ELECTING

THE

MEASU

-

RING

METHOD

, page 56).

2.

Prepare two standard calibration solutions.

3.

Perform a two-point calibration according to user guidance.

4.

The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.

01.08.2013 08:00

01.08.2013 08:00

#1 0.010 mg/l

20.2 mV 25.0 °C

#2 0.020 mg/l

79.2 mV 25.0 °C

Slp.: 54.1 mV
Sensor +++ (Ion type: Ag)

CALIBRATION
ADA 94/IDS BNC
B092500013

5.

Start the measurement with <F1>/[continue].
An entry window appears.

6.

Prepare the sample.

7.

If necessary, add the ISA/TISAB solution to the sample.

8.

Thoroughly rinse the electrode with deionized water.

9.

Immerse the electrode in the sample.

10.

Using <> <> and <ENTER>, select the values for the volume of
the sample (Sample volume) and the volume of the ISA/TISAB solution
(ISA/TISAB vol.).

i

Immerse sensor in sample

Sample volume

100.0 ml

ISA/TISAB vol.

1.0 ml

continue

Standard addition

01.08.2013 08:00