YSI MultiLab 4010-3 User Manual

Page 67

Advertising
background image

MultiLab 4010-3

Ion concentration

ba76143e03

08/2013

67

Measuring

01.08.2013 08:00

01.08.2013 08:00

#1 0.010 mg/l

20.2 mV 25.0 °C

#2 0.020 mg/l

79.2 mV 25.0 °C

Slp.: 54.1 mV
Sensor +++ (Ion type: Ag)

CALIBRATION
ADA 94/IDS BNC
B092500013

5.

Start the measurement with <F1> continue
An entry window appears.

6.

Prepare the sample.

7.

If necessary, add the ISA/TISAB solution to the sample.

8.

Thoroughly rinse the electrode with deionized water.

9.

Immerse the electrode in the sample that was supplemented with blank
value solution.

10.

Using <> <> and <ENTER>, enter the values for the volume of the
sample (Sample volume), the volume of the ISA/TISAB solution (ISA/
TISAB vol.
), the volume of the blank value solution (BV volume) and the
concentration of the blank value solution (BV conc.).

11.

Select continue with <> <> and start the measurement with
<ENTER>.
When the measurement is finished an entry window appears.

i

Immerse sensor in sample

Sample volume

100.0 ml

ISA/TISAB vol.

1.0 ml

BV volume

100.0 ml

BV conc.

1.000 mg/l

continue

Blank value addition

01.08.2013 08:00