YSI MultiLab 4010-3 User Manual

Page 65

Advertising
background image

MultiLab 4010-3

Ion concentration

ba76143e03

08/2013

65

Measuring

7.

Start the measurement with <F1> continue.
An entry window appears.

8.

Prepare the standard solution.

9.

If necessary, add the ISA/TISAB solution to the standard solution.

10.

Thoroughly rinse the electrode with deionized water.

11.

Immerse the electrode in the standard solution (complexing agent or
precipitating agent).

12.

Using <> <> and <ENTER>, enter the values for the volume of the
standard solution (Std. volume), the concentration of the standard solu-
tion (Std. conc.) and the volume of the ISA/TISAB solution (ISA/TISAB
vol.
).

13.

Select continue with <> <> and start the measurement with
<ENTER>.
When the measurement is finished an entry window appears.

14.

Add the sample to the standard (complexing agent or precipitating
agent).

15.

Using <> <> and <ENTER>, enter the value for the volume of the
sample (Sample volume).

16.

Select continue with <> <> and start the measurement with
<ENTER>.
The measurement result is displayed when the measurement is com-
pleted.

i

Immerse sensor in standard

Std. volume

100.0 ml

Std. conc.

1.0 mg/l

ISA/TISAB vol.

1.0 ml

continue

Sample subtraction

01.08.2013 08:00

i

Add sample!

Sample volume

1.0 ml

continue

Sample subtraction

01.08.2013 08:00