Dissolved oxygen (D.O.)

MultiLab 4010-3

76

ba76143e03

08/2013

Calibration

evaluation

After calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.

ProOBOD

calibration

evaluation

Menu item

Setting/func-
tion

Explanation

Calibration /
Calibration data
storage
/Display

-

Displays the calibration records.
Further options:

Scroll through the calibration
records with <><>.

Using <PRT>, output the dis-
played calibration record to
the USB-B (PC) interface or
the USB-A (USB printer)
interface .

Using <PRT_>, output all cal-
ibration records to the USB-B
(PC) interface or the USB-A
(USB printer) interface .

Quit the display with <ESC>.

Switch directly to the mea-
sured value display with <M>.

Calibration / Cali-
bration data stor-
age
/ Output to
USB flash drive or
printer

-

Outputs the stored calibration
data to the USB-A interface
(USB memory device/USB
printer)

Calibration /
Calibration data
storage
/ Output to
RS232/USB

-

Outputs the stored calibration
data to the USB-B interface (PC)

Display

Calibration record

Relative slope

+++

S = 0.94 ... 1.06

++

S = 0.92 ... 0.94
or
S = 1.06 ... 1.08

+

S = 0.90 ... 0.92
or
S = 1.08 ... 1.10

Error

Error

S < 0.90
or
S > 1.10

Error elimination (see section 14 W

HAT

TO

DO

IF

...,

page 115)