10.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.

After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC

FUNCTION

(

SENSOR

QUALITY

CONTROL

), page

40).

After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.

Setting the calibra-

tion interval

The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):

22

Reagecon 20 *

2.000
4.000
7.000
10.000
13.000

20 °C

23

Reagecon 25 *

2.000
4.000
7.000
10.000
13.000

25 °C

24

Chemsolute *

2.000
4.000
7.000
10.000

20 °C

25

USABlueBook

*

4.000
7.000
10.000

25 °C

The buffers are selected in the menu, pH / <ENTER> / Calibration
/ Buffer (see section 10.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

, page

84).

To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.

No.

Buffer set *

pH values

at

1.

Open the menu for measurement settings with <ENTER>.

2.

In the Calibration / Calibration interval menu, set the calibration interval
with <><>.