YSI MultiLab 4010-3 User Manual

Page 93

Advertising
background image

MultiLab 4010-3

Settings

ba76143e03

08/2013

93

Calibration / Calibra-
tion data storage
/
Display

-

Displays the last calibration records
(max. 10)

Calibration / Calibra-
tion data storage
/
Output to USB flash
drive or printer

-

Outputs the stored calibration data to
the USB-A interface (USB memory
device/USB printer)

Calibration /
Calibration data stor-
age
/ Output to
RS232/USB

-

Outputs the stored calibration data to
the USB-B interface (PC)

Calibration /Calibra-
tion interval

1 ... 150 ...
999 d

Calibration interval for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.

Type

Used measuring cell

Cal

Measuring cells whose cell constant
is determined by calibration in the KCl
control standard solution.
Calibration range:
0.450 to 0.500 cm

-1

The currently valid cell constant is dis-
played in the status line.

man

Freely adjustable cell constant in the
range 0.450 ... 0.500 cm

-1

.

Man. cell const.

0.450 ...
0.475 ...
0.500 cm

-1

Display and setting options for the cell
constant.

Temp. comp. (TC) /
Method

nLF
lin
off

Procedure for temperature compen-
sation (see section 9.2 T

EMPERATURE

COMPENSATION

, page 80).

This setting is only available for the
measured parameters,

ϰ and ρ.

Temp. comp. (TC) /
Linear coeff.

0.000 ...
2.000 ...
3.000 %/K

Coefficient of the linear temperature
compensation.
This menu item is only available when
the linear temperature compensation
is set.

Temp. comp. (TC) /
Reference temp.

20 °C
25 °C

Reference temperature
This setting is only available for the
measured parameters,

ϰ and ρ.

Multiplier for TDS

0.40 ... 1.00

Factor for TDS value

Menu item

Possible
setting

Explanation