MultiLab 4010-3

pH value

ba76143e03

08/2013

39

The following information is displayed:

The limits of the calibrated range are determined by the buffers used for cali-
bration:

1 Currently measured pH value (needle)
2 Marking lines for all nominal buffer values used with the last valid cali-

bration

3 Measuring range for which a valid calibration is available. Measured

values in this range are suitable for documentation.

4 Measuring range for which no valid calibration is available (dark

gray). Measured values in this range are not suitable for documenta-
tion. Calibrate the meter with buffers covering this measuring range.
If the current measured value is outside the calibrated range, this area
is displayed in a darker gray.
If a measured value is outside the measuring range pH 0 - 14, over-
flow arrows are displayed at the left or right edge of the measuring
range.

Lower limit:

Buffer with lowest pH value - 2 pH units

Upper limit:

Buffer with highest pH value + 2 pH units

1
2

3

4