9 Indicating transmitter specifications

97

9.7 Display messages

Message

Cause and action

NORMAL OPERATION

No faults.

DIVERT CONTROL

Not used with PR-03.
If this message occurs, deactivate with the command
Calibrate / Parameters / Divert control /

Divert Control

.

REMOVED FROM DIVERT CONTROL

Not used with PR-03

RETRYING WASH

The prism wash failed the wash check (Section 5.9.4). The

(Sanitary refractometer PR-03-A

prism wash will be attempted as many times as is set by the

and probe refractometer PR-03-P)

Wash retry

parameter (Section 5.9.4).

WASH STOP/TEMP LIMIT

The process temperature is below the temperature limit set for

(Sanitary refractometer PR-03-A

wash stop (Section 5.9.3). The message is shown until wash

and Probe refractometer PR-03-P)

succeeds or the RESET key is pushed.

LOW CONCENTRATION/NO SAMPLE

Normally caused by empty process pipe or exceptionally low
concentration.
If pipe is full and sample indicates normal concentration,
check the optical image (Section 7.2).

HIGH PROCESS TEMPERATURE

Process temperature is above the set limit (Section 5.7). Com-
mon causes are abnormal process conditions or a leaking
prism wash steam valve.

LOW LIGHT INTENSITY

Most common cause is onset of prism coating.

DIP IN IMAGE

Indicates irregular optical image. Can be caused by a speck of
dust on the CCD window.

HIGH CONC/PRISM COATED

Is most commonly caused by prism coating, but may also indi-
cate exceptionally high concentration. If sample is normal and
the message persists after the prism surface is cleaned, check
the optical image (Section 7.2).

** WASH STOP/EXT. STOP **

Caused by closing an input switch that has been activated for

(Sanitary refractometer PR-03-A

external wash stop (Section 3.2.2.6 and Section 5.6).

and probe refractometer PR-03-P)