5 Startup, configuration and calibration adjustment

35

HT

Sensor head temperature sensor reading. This value is used in calculating the actual Sensor
head internal temperature.

HH

Sensor head humidity sensor reading. The actual relative humidity inside the sensor is cal-
culated from HH and HT.

5.5.3 Viewing Scaled image

In the scaled image (Figure 5.8) the optical image has been modified mathematically to make a clearer
curve.

A

B

C

D

SCALED

OPTICAL

IMAGE

TEST: 115.9

51.2

°C

Endp: 21

L):

9.8

R):

25.9

HT: 20 °C

HH: 3%

Slope

Normal operation

Figure 5.8

Scaled image display

The Scaled image display provides following information:

TEST:

The calculated TEST value, i.e. the number of photocells on the lisght side of the optical image
(to the left fo the vertical line). Range 8–248. Indicated by a dotted vertical line.

(T)

Indicates the process temperature.

Endp:

The vertical intensity value of the rightmost point in % of span.

L):

A measure of the left side curvature of the optical image.

R(:

A measure of the right side curvature of the optical image.
A zero curvature indicates a straight line. Both curvature values are defined as positive for an
image as Figure 5.5.3.

HT:

Sensor head internal temperature in

C.

HH:

Sensor head internal relative humidity in %RH.

5.5.4 Viewing slope

The Slope display (Figure 5.9) shows the slope curve of the optical image. Slope is a measure of the
sharpness of the optical image.

The SLOPE curve should have a sharp dip at the TEST value. The following additional values are given: