7 configuring relays, Ab c d – K-Patents PR-03 User Manual

Page 43

Advertising
background image

43

43

43

43

5 Startup, configuration and calibration adjustment

37

A

B

C

D

RI(25ºC

1.4412

SWITCH INPUT A (Not defined)

0 Not defined

1 Scale select

2 Ext hold & wash start

3 External wash stop

4 Calibration seal

Switch

B

Switch

C

Switch

D

Figure 5.10

Switch input display for input A

D).

Note:

The range will not change.

Example:

medium 1 = 20 - 40 % Sugar, medium 2 = 20 - 40 % Salt.

2

Ext hold & wash start

Remote wash start for Relay unit wash function and external Hold.
This switch function is useful for an intermittent process: the prism
is washed when the process stops and again when it starts, between
the two washes the output signal is on Hold.
Only Sanitary refractometer PR-03-A, Probe refractometer PR-03-P
and Process refractometer PR-03-D

.

3

External wash stop

Only Sanitary refractometer PR-03-A, Probe refractometer PR-03-P
and Process refractometer PR-03-D

.

4

Calibration seal

An input switch can be configured for calibration seal. Activating and
connecting the input switch for calibration seal will prevent calibra-
tion through software.

5.7 Configuring relays

You can check the existing relay configuration and functions with the Indicating transmitter. Press the
key sequence Display/System configuration/Relay configuration to display the current
relay settings.

The default setting for built-in signal relay 1 is No malfunction and for signal relay 2 Internal
humidity above 50 %

. A closed contact on the Relay 1 indicates that the instrument works properly.

It is recommended to use this relay for alarm purpose in a control system.

In relay units, Relay B is configured for wash by default.

You can reprogram the relay functions any time through the IT-R’s calibration menu by performing the
following steps:

Advertising