Ab c d – K-Patents PR-03 User Manual

Page 38

Advertising
background image

38

38

38

38

32

PR-03 instruction manual

Example:

You need to change the bias (see Section 5.10.2) to be 25.456 (as in the example in

Section 5.10.2). Starting from the Normal display (Section 5.3), you have to perform following
steps:

1. Give the right command sequence to get to the value to be changed, in this case

Calibrate/Parameters/CONC (R.I.)/Parameters/Bias
(key sequence C-B-A-A-1).

2. Type the new value.
3. When the new value is finished, press ENTER.
4. Accept (=reconfirm) the new value by pressing ENTER again. The new value now appears on

the screen as the current value.

5. Finally, withdraw from the current display by pressing the RESET key as many times as neces-

sary (five times to get to the Normal display from the Bias setting).

5.4 Soft key Display: Getting information on the process

and the settings

Pressing the soft key Display (key D) in the Normal display brings up the Information display , Fig-
ure 5.5 below. Through the Information display you can get information about the process, the process
parameters and other settings that have been entered to the Indicating transmitter. This branch of the se-
lection tree (Figure 9.5) is safe, because it doesn’t allow you to make changes to the settings, it will only
display the existing information.

A

B

C

D

CONC

36.7%

TEMPERATURE: 30.2 °C

86.4 °F

STANDARD RI(25°C): 1.3960

TEST: 133.6

Normal operation

Optical

image

System

configur-

ation

Sensor

head

Output signal: 16.7 mA

Figure 5.5

The Information display

As you can see in Figure 5.5, the Information display contains additional data compared to the Normal
display:

− The process temperature in both

C and

F

− The standard RI (25

C). This shows the Refractive Index of liquid applied to the prism, referenced to

25

C

− output current in mA

Advertising