5 Startup, configuration and calibration adjustment

39

5.8 Configuring external output unit

To program the External output unit from the keyboard of the Indicating transmitter, start with the sequence
Calibrate / Parameters / Output signals / External output

to get the External

output unit display (Figure 5.11).

A

B

C

D

RI(25ºC

1.4412

EXTERNAL OUTPUT UNIT (Not connected)

1 Zero (0.0000)
2 Span (0.0000)
3 Hold function (Active)
4 Range (4..20 mA)
5 Source (Not defined)

Figure 5.11

External output unit display

1

Zero

Measurement value corresponding to 4 mA or 0 mA (depending on your cho-
sen range)

2

Span

Measurement value span corresponding to mA output span

3

Hold function

Select either 0 Inactive or 1 Active

4

Range

Select either 0 4..20 mA or 1 0..20 mA

5

Source

Select from
0 Not defined

1 RI (25

C)

2 Standard RI (25

C)

3 Temp

C

4 Temp

F

5 TEST

6 Sensor temperature

7 Sensor humidity

Example:

For source

C, zero = 20, span = 80, the output signal 4 - 20 mA corresponds to 20-100

C.

5.9 Configuring automatic prism wash

The relay units -WR and PR-7080 can be configured to control prism wash. Automatic prism wash can be
initiated with regular intervals (Section 5.9.1) or based on certain measurement values (Section 5.9.2). It
is also possible to configure these automatic wash functions so that they are not used when the process is
stopped or the pipe is empty (Section 5.9.3 and Section 5.9.4).