10 PR-03 process refractometers in potentially explosive atmosphere

101

10 PR-03 process refractometers in potentially explosive

atmosphere

The PR-03 refractometer series can be used in locations with potentially explosive atmosphere with follow-
ing modifications, made by K-Patents Oy

. The refractometer sensor’s compliance with the Essential Health

and Safety Requirements are assured by complying with standard EN 50 021:1999.

The PR-03-...-AX refractometers have been certified by KEMA Quality B.V. under European ATEX direc-
tive 94/9/EC for ATEX EX II 3 G/Cenelec Eex nA IIC T4. The EX-Type Certificate Number is KEMA
04ATEX 1088X. The report number is KEMa No. 207 1534.

10.1 Equipment

The K-Patents refractometer system (Figure 10.1) for potentially explosive atmosphere locations consists of
a modified refractometer sensor PR-03-...-AX, a standard Indicating transmitter IT-R and a standard cable
PR-8300-...

Hazardous area

(up to Zone 2)

Safe Area

Figure 10.1

Refractometer system PR-03-...-AX