PR-03 instruction manual

8.4.5 PR-03-P mounting specifics

The Probe refractometer PR-03-P is primarily designed for mounting in a tank wall. To ensure that the
measurement is representative and that the prism stays clean, the probe should be installed close to a stirrer,
Figure 8.8.

4-20 mA

~

CHECK CABLE LENGTH

Figure 8.8

Recommendations for mounting the Probe sensor PR-03-P in a tank

K-Patents probe refractometer type PR-03-P62-TSS is connected to the process by a 2 1/2" 3-A sanitary
clamp, Figure 8.9. The type PR-03-P62-RSS type sensor is connected to process by a 4" 3-A sanitary
clamp.

Note:

For higher process (or ambient) temperatures, use instead a flush mounted sensor (Figures 8.9, 8.10

and 8.11), where the electronics in the sensor head are further away from the process heat.

Figure 8.9

Insertion of Probe refractometer PR-03-P62-TSS