PR-03 instruction manual

IMAGE DIAGNOSTICS

Max intensity OK (above 100)?
Endpoint below 75%?
Image below corner?
Slope OK (above 1)?
Left curve L) OK (above 1)?
Right curve R) OK (above 1)?

Yes
Yes
Yes
Yes
Yes
Yes

SENSOR HEAD

Head temperature: 28

°C

Head humidity: 3%

Normal operation.

A

B

C

D

CONC

68.0%

PROCESS TEMPERATURE: 31.2 °C
STANDARD RI(25°C): 1.4194

TEST: 115.7
Normal operation

Calibrate

Display

A

B

C

D

A

B

C

D

CONC

PROCESS TEMPERATURE: 32.5

°C

STANDARD RI(25

°C): 1.4190

TEST: 117.2
Normal operation

Optical

image

Para-

meters

68.0%

A

B

C

D

RAW
SENSOR
DATA

RMN: 2
RMX: 216
LED: 96
Scans: 2
A/D: 314
Sts: 00h
HT: 143
HH: 10

Scaled
image

Normal operation

A

B

C

D

SCALED
OPTICAL
IMAGE

TEST: 115.9
51.2

°C

Endp: 21
L(: 9.8
R): 25.9
HT: 20 °C
HH: 3%

Slope

Normal operation

Diagnos-
tics

SLOPE

TEST: 115.7
31.2

°C

Max: 208
Slope: 2.7
Endp: 21
L): 10.1
R(: 25.8

Normal operation

A

B

C

D

Good LED value 20-150

Good slope value
more than 1.5

Good HT lower than 60 ºC
Good HH lower than 10 %

Figure 7.5

Troubleshooting your refractometer system with IT-R