PR-03 instruction manual

1. Take the sensor from the process and place it on a table.

Note:

Keep the PR-03-M with the flow cell and place it on a table with the flow cell (medium adaptor)

up. Use the flow cell as a sample holder.

2. Allow the sensor temperature to settle close to 25

C. The temperature does not have to be exactly 25

C,

since the differences in the calibration temperature are compensated by the software.

3. Ensure that the prism is clean without any traces of the process liquid. Use a cleaning solvent appropriate

for the process medium.

4. Use the certified standard R.I. liquids, which correspond to the values indicated in the Final test section

of the latest Delivery Data Sheet or Standard calibration sheet of your PR-03 sensor. The three R.I.
values are selected by K-Patents calibration software to represent the used R.I. range.

5. Check the R.I. calibration with the three R.I. liquids. The use of K-Patents PR-03 sample holder (see

Figure 5.15) or of the flow cell (medium adaptor) is recommended to ensure a right portion of liquid on
the prism.

6. The R.I. calibration can be checked against the standard R.I. liquid values as follows:

Display

Key-in sequence

Value to be checked

Acceptable value

from Normal display

Information display

D

Standard RI (25

C)

Sample R.I. ± 0.0006

Corrective action if the verification fails

If the verification does not meet the acceptance criteria defined above, the following steps can be taken to
correct the calibration of PR-03.

1. Ensure that the prism is in good condition

See Section 6 for more details of prism mounting and dismounting.

2. The following values and diagnostics should be checked from the Indicating transmitter displays with

the standard R.I. liquid on the prism.

Display

Key-in sequence

Value to be checked

Acceptable value

from Normal Display

Raw sensor data

C-A

LED

below 200

Image diagnostics

C-A-A-A-A

Max intensity OK

Yes

Endpoint below 75%

Yes

Image below corner

Yes

Slope OK

Yes

Left curve OK

Yes

Right curve OK

Yes