4 alarms menu – ONICON F-2600 User Manual

Page 61

Advertising
background image

11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]

F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204

Page 61

3.3.4 Alarms Menu

Password

ENTER

Use

keys to access menus

Alarms

Menu

Relay Alarm 1

More >

ENTER

Run Mode

Relay Alarm 2

More >

Relay Alarm 3

More >

Alarm LOG

xx Files (ENTER)

Clear Alarm LOG?

YES or NO

<Measure> units

xxxx

xxxx

xxxx

Alarm File

Time

Date

(Press EXIT to return

to Alarm LOG.)

* see below

Physical Layer does not exist on Two Wire

Mass - Accessible via HART

<Measure>

None

Mass

Volume

Temp 1,2

Press

Density

** Energy

<Measure>

None

Mass

Volume

Temp 1,2

Press

Density

** Energy

<Measure>

None

Mass

Volume

Temp 1,2

Press

Density

** Energy

<Mode>

None

HIGH Alarm (>)

LOW Alarm (<)

<Mode>

None

HIGH Alarm (>)

LOW Alarm (<)

<Mode>

None

HIGH Alarm (>)

LOW Alarm (<)

<Measure> units

<Measure> units

*

**

Energy Meters Only

Example for Setting an Alarm

The following shows how to set Relay Alarm 1 to activate if the mass flow rate is greater
than 100 lb/hr. You can check the alarm configuration in the Run Mode by pressing the

⇑ ⇓ keys until Alarm (1) appears. The lower line displays the mass flow rate at which the
alarm activates. Note: All outputs are disabled while using the Set-up Menus.

First, set the desired units of measurement:

1. Use ⇐ ⇒ keys to move to the Units Menu (see to page 64).
2. Press ⇓ key until Mass Flow Unit appears. Press ENTER.
3. Press ⇓ key until lb appears in the numerator. Press ⇒ key to move the underline

cursor to the denominator. Press the ⇓ key until hr appears in the denominator. Press
ENTER to select.

4. Press ⇑ key until Units Menu appears.

Second, set the alarm:

1. Use ⇐ ⇒ keys to move to the Alarms Menu.
2. Press the ⇓ key until Relay Alarm 1 appears.
3. Press ⇒ key to access Measure selections. Press ENTER and use the ⇓ key to select

Mass. Press ENTER.

4. Press ⇒ key to select the alarm Mode. Press ENTER and use ⇓ key to select HIGH

Alarm. Press ENTER.

5. Press ⇒ key to select the value that must be exceeded before the alarm activates.

Press ENTER and use ⇑ ⇓ ⇐ ⇒ keys to set 100 or 100.0. Press ENTER.

6. Press the EXIT key to save your changes. (Alarm changes are always permanently

saved. Up to three relay alarm outputs are available depending on meter
configuration.)

Advertising
This manual is related to the following products: