M4557 - Model 9210 Nitriding Controller

SSi Manual SERIES 9210-M4557-Nitriding

Page 54 of 54

DO NOT CHANGE THE FILTER WHILE AMMONIA OR DISASSOCIATED AMMONIA IS FLOWING.

THE GASES ARE TOXIC AND VERY HAZARDOUS.


The steps for replacing the filter are as follows;

1. Close all incoming gas valves.

2. Unscrew the clear bowl from the filter housing.

3. Unscrew the filter holder. The holder is the black plastic knob in the center of the filter.

4. Pull off old filter and replace with new.

5. Screw the filter holder back into the top of the filter housing.
6. Screw the clear bowl back on the housing.

7. Open incoming gas lines

Flow to the sample cell should be a minimum of .5 scfh and a maximum of 2.0. This flow is critical to the

nitriding process. If flow is low replace the filter. If you are unable to maintain flow contact Super

Systems Inc at 1-800-666-4330 immediately.