M4557 - Model 9210 Nitriding Controller

SSi Manual SERIES 9210-M4557-Nitriding

Page 61 of 61

Quantity: is one of H2 CO CO2

xxxx: is the measured quantity. The resolution is the same as the display:
Unit: is one of % percent r compensation ratio

The compensation ratio is the diagnostic data for instruments with a secondary gas background. When

requested as a data it will return the present ratio for compensating the primary measured gas “R1”

I.e. for H2 in N2 with a secondary gas of CO2

D1 CO2=0.069r If you know the concentration of the CO2 can calculate the correction in the H2 reading as

H2- ( CO2 * 0.069)
In over-range condition (when the display shows ‘HI’) xxxxx will be ‘+++++’ In under-range condition

(when the display shows ‘LO’) xxxxx will be ‘-----’

Controller.

Sensor

Reading

R1 H2= 98.5%

Instrument request a
reading

Span=99.0

S1 Pass

Controller requests a span
setting of 99.0%

Reading

R1 H2= 99.0%

I.e. this would set the present instrument reading to 99.0% The unit “%” is assumed to be the same as the
reading.

Error messages
Error messages take the form ‘? xx’ where ‘xx’ is a numeric code as explained below Comms. Errors:

Code

Error

Description

90

Buffer overflow

More than 15 characters were received without message terminator
(<CR><LF>). Any subsequent characters will begin a new message.

91

Message timeout

10 seconds has elapsed since the last character was received
without message terminator.

92

Bad opcode

Message was received correctly terminated but not understood (e.g.
‘Fred=1<CR><LF>’)

93

Bad operand

Message was received correctly terminated and understood but the
line number wasn’t (e.g. ‘Reading=Q<CR><LF>’).