Ethernet interface

Page 5

- Decimal digital data:

Made of a succession of ASCII-coded characters beginning with a digit or a sign (+ or -) It can be
of NR1 (integer), NR2 (decimal) or NR3 type (with an exponent).
For example for a digital character: -25.02.

- Text data:

7-bit ASCII-coded character chain between quotation marks (") or apostrophes (').
For example: "A".

- Any succession ASCII characters: ends with the termination sequence.

14.3.

Standard instructions

All these instructions begin with "*".

*IDN ?

REQUEST FOR IDENTIFICATION OF AN APPLIANCE

answer by the appliance: 4 pieces of data separated by ',':

-

the trademark of the appliance

-

the name of the appliance, followed by _nn where nn is the number of inputs of the
recorder

-

the serial number of the appliance (0 if unknown)

-

the version number of the software: x.xx x

*OPT ?

REQUEST FOR IDENTIFICATION OF THE OPTIONS OF AN APPLIANCE

answer by the appliance: n pieces of data separated by ',':

-

number of input modules

-

number of channels per module

*RST

RESET OF AN APPLIANCE

action: initialisation of the recorder in a fixed configuration (voltage inputs, 10V range, 0V centre...)

*REM

SWITCHING TO PROGRAMMATION (REMOTE)

required in RS232C before sending any other programming command.

*LOC

SWITCHING TO LOCAL MODE

*CLS

DELETING THE STATUS REGISTERS

action: the appliance resets the status registers to zero.

*ESE

VALIDATION OF THE STANDART EVENT BITS OF AN APPLIANCE

*ESE is followed with a digit, from 0 up to 255

action: changes the validation register of the standard events and updates the ESB bit in the status

register of the service request (see the following paragraph).

*ESE ?

INTERROGATION OF THE CONTENT OF THE VALIDATION REGISTER OF

THE STANDARD EVENTS OF AN APPLIANCE

answer by the appliance: NR1 number between 0 and 255 (see the following paragraph).