16 Technical specifications

Page 3

16.2. Multiplexed inputs

16.2.1.General characteristics

Number of inputs per module: 12

Inputs type differential non isolated

Impedance:

Impedance > 10 M

 for ranges  2 V

Impedance = 2 M

 for any other range

Maximum possible voltages:

Between a measure channel and the ground:

48 V DC

Between the 2 terminals of a channel:

48 V DC

Maximal common mode voltages :

3 V for ranges

 2 V

50 V for any other range

Type of measurements:

Voltage, current (by external shunt)
Thermocouple J, K, T, S, B, N, E, W5
PT100 2, 3 or 4 wires

16.2.2.Voltage recording

Maximum range

50 V

Minimum range

1 mV

Offset

Adjustment of the centre by software

Accuracy

0.1% of the full scale, 10 V, 0.1% of the offset

Offset drift

100 ppm/ C 1 V/ C

16.2.3.RMS recording

RMS calculation by software

Resolution

200 s

Maximum frequency

500 Hz

Crest factor

2.2

Accuracy

1 % (wave signal)

Response time

typical. 100 ms