16 Technical specifications
Page 1
16. TECHNICAL SPECIFICATIONS
16.1. Isolated inputs
16.1.1.General characteristics
Number of inputs per module: 6
Impedance:
Impedance > 25 M
for ranges 1 V
Impedance = 1 M
for any other range
Maximum possible voltages:
Between a measure channel and the ground:
+ 500 V DC or 500 V AC 50 Hz
Between the 2 terminals of a channel:
+500 V DC or 500 V AC 50 Hz
Category of installation: overvoltage category: III 600V
Isolation: between the ground and the measure channel: >100 M at 500 V DC.
Spurious common mode voltages: test according to EN 61143
Types of measures :
Voltage, current (by external shunt)
Frequency
Thermocouple probe J, K, T, S, B, N, E, W5
16.1.2.Voltage recording
Maximum range
1000 V
(-500 V to +500 V)
Minimum range
1 mV
(-0.5 mV to +0.5 mV)
Offset
Adjustment of the centre by 1/5000 of the full scale or by -range
Maximum offset
+ 5 range (except 1000V)
Accuracy
0.1% of the full scale, 10 V, 0.1% of the zero offset
Offset drift
100 ppm/ C, 1 V/ C
Class C index
see appendix
16.1.3.RMS recording
RMS calculation by the software
Resolution
200 s
Maximum frequency
500 Hz
Crest factor
2.2 and 600 V max instant.
Accuracy
1 % (wave signal)
Response time
typical. 100 ms
Measurable max. voltage
424 V AC