Error messages, Line on. the last line is followed by a blank line – Escort 3016B User Manual
Page 118
15
Ethernet interface
Page 20
14.5.15.
Service request
See the explanations about the structure of status data.
14.6.
Error messages
When a problem arouses in the programming via the interface of the recorder, a Debug window is
displayed on screen to help you identify your error:
Each error is associated with a line that shows:
-
an error number
-
the received message
When the window is full, the errors are displayed again from the 1
st
line on.
The last line is followed by a blank line.
HEADER
PARAMETERS
EXAMPLES
SRQ_ENABLE
P1
SRQ_ENABLE 3
Allows to modify the alarm enable register
P1 = register value
3 = 1 + 2 (bits 0 and 1)
Start and End of plot events are
authorized to be indicated in the status
byte register (bit 0).
bit
decimale value
use
0
1
start of plot
1
2
end of plot
2
4
end of write
3
8
no more paper
4
16
door open
5
32
start of memory record
6
64
end of memory record
7
128 trigger of memory record
SRQ_ENABLE ?
Returns the value of the alarm enable register
SRQ_TYPE ?
Returns the value of the alarm status register.
SRQ_TYPE ?
The register is cleared.
the recorder returns : SRQ_TYPE 4
Each bit of this value is identical to the previously defined bit in
SRQ_ENABLE.
so "a write operation is completed"
Error N
Explanation
1
Unknown header
2
Unknown parameter
3
Prohibited parameter
4
Missing parameter
5
Incorrect parameter separator
6
Incorrect message separator
7
Too long word
8
Incorrect format of text parameter
9
Prohibited query
10
Numerical parameter out of limits
11
Text parameter out of limits
12
Obligatory query
13
Transmission buffer full
14
Not possible in this contexte
15
Checksum error