Escort 3016B User Manual
Page 121
16 Technical specifications
Page 2
16.1.4.Temperature recording
CAPTEUR
using environment
Ranges
PT100
COUPLE J
CO UPLE K
COUPLE T
CO UPLE S
CO UPLE B
CO UPLE E
CO UPLE N
COUPLE W5
-200 C to 850 C
-210 C to 1200 C
-250 C to 1370 C
-200 C to 400 C
-50 C to 1760 C
200 C to 1820 C
-250 C to 1000 C
-250 C to 1300 C
0 C to 2320 C
20 C to 1000 C
20 C to 2000 C
20 C to 2000 C
20 C to 500 C
50 C to 2000 C
50 C to 2000 C
20 C to 1000 C
20 C to 1000 C
50 C to 2000 C
The accuracy of the thermocouples is indicated in the appendix
Compensation of the cold junction of the thermocouples J,K,T,S,N,E,W5: 1.25 C
16.1.5.Sampling
Resolution:
14 bits
Max. sampling period:
Memory mode:
1 s (i.e. 1 MHz)
Direct mode:
200 s (i.e. 5 kHz)
Max. sampling duration:
10 min
16.1.6. Bandwidth
Bandwith at -3 dB:
Range
Bandwidth
> 1 V
100 kHz
> 50 mV
50 kHz
20 mV
30 kHz
10 mV
30 kHz
5 mV
20 kHz
Internal analogical filters:
10 kHz ,1 kHz, 100 Hz, 10Hz
slope:
20 dB/decade
Software filters:
1 Hz, 0,1 Hz, 0,01 Hz, 0,001 Hz
Frequency record:
Sensitivity
100 mVrms min.
Minimum duty factor
10 %.
Frequency
between 10 Hz and 100 kHz.
Accuracy
0.2 % of the full scale