Escort 3016B User Manual

Page 35

Advertising
background image

4

Use

Page 6

4.5.

Channels set up key

Configuration of the channels. After pressing this key, select a module, then select a channel to
access its parameters.

4.5.1. Analogical channels

Name: give a name to the channel (max. 26 characters)

Type: choice of the type of measure made on this input

-

voltage, intensity, frequency or thermocouple

-

direct or RMS

-

value of the shunt in intensity measurement

-

choice of the type in thermocouple, compensation, unit

Filter: positioning a filter at the inlet

-

10 kHz, 1 kHz, 100 Hz, 10 Hz for analogical filters

-

1 Hz, 10 s, 100 s or 1000 s for digital filters

Function: makes it possible to assign a mathematical calculation function to the considered

channel

-

Without: no function.

-

Change unit: modifies the unit of the measures on the channel; you can then
program a couple of dots X1, Y1 and X2, Y2 to achieve a scaling

-

Calculation: available mathematical functions, associated parameters and unit

Advertising